ePrints@IIScePrints@IISc Home | About | Browse | Latest Additions | Advanced Search | Contact | Help

High-pressure low-temperature locknut cell for both EPR and NMR studies to 10 kilobars and 77 K

Sinha, Shantanu and Srinivasan, R (1983) High-pressure low-temperature locknut cell for both EPR and NMR studies to 10 kilobars and 77 K. In: Review of Scientific Instruments, 54 (11). pp. 1492-1496.

[img] PDF
6.pdf
Restricted to Registered users only

Download (462kB) | Request a copy

Abstract

A locked high-pressure cell with working pressure range up to 10 kbars suitable for low-temperature studies to 77 K has been described. It can be used for both EPR and NMR studies of single crystals (and other solid samples). The high-pressure seal and all other aspects of the cell remain the same for either application. Only a change of the bottom plug is required for a switch from a nuclear-magnetic-resonance (NMR) to an electron-paramagnetic-resonance (EPR) experiment. Details of the procedure for the calibration of pressure inside the cell at various temperatures are discussed. The performance of the cell in EPR ($Cr^{3+}ion$) and NMR ($^{27}Al$ nucleus) studies is reported.

Item Type: Journal Article
Publication: Review of Scientific Instruments
Publisher: American Institute of Physics
Additional Information: Copyright of this article belongs to American Institute of Physics.
Department/Centre: Division of Physical & Mathematical Sciences > Physics
Date Deposited: 14 Jul 2008
Last Modified: 19 Sep 2010 04:47
URI: http://eprints.iisc.ac.in/id/eprint/15008

Actions (login required)

View Item View Item