Vinayakumar, KB and Kumar, Naveen G and Nayak, MM and Dinesh, NS and Rajanna, K (2015) Peristaltic pump-based low range pressure sensor calibration system. In: REVIEW OF SCIENTIFIC INSTRUMENTS, 86 (11).
PDF
Rev_Sci_Ins_86-11_115113_2015.pdf - Published Version Restricted to Registered users only Download (7MB) | Request a copy |
Abstract
Peristaltic pumps were normally used to pump liquids in several chemical and biological applications. In the present study, a peristaltic pump was used to pressurize the chamber (positive as well negative pressures) using atmospheric air. In the present paper, we discuss the development and performance study of an automatic pressurization system to calibrate low range (millibar) pressure sensors. The system includes a peristaltic pump, calibrated pressure sensor (master sensor), pressure chamber, and the control electronics. An in-house developed peristaltic pump was used to pressurize the chamber. A closed loop control system has been developed to detect and adjust the pressure leaks in the chamber. The complete system has been integrated into a portable product. The system performance has been studied for a step response and steady state errors. The system is portable, free from oil contaminants, and consumes less power compared to existing pressure calibration systems. The veracity of the system was verified by calibrating an unknown diaphragm based pressure sensor and the results obtained were satisfactory. (C) 2015 AIP Publishing LLC.
Item Type: | Journal Article |
---|---|
Publication: | REVIEW OF SCIENTIFIC INSTRUMENTS |
Publisher: | AMER INST PHYSICS |
Additional Information: | Copy right for this article belongs to the AMER INST PHYSICS, 1305 WALT WHITMAN RD, STE 300, MELVILLE, NY 11747-4501 USA |
Department/Centre: | Division of Electrical Sciences > Electronic Systems Engineering (Formerly Centre for Electronic Design & Technology) Division of Interdisciplinary Sciences > Centre for Nano Science and Engineering Division of Physical & Mathematical Sciences > Instrumentation Appiled Physics |
Date Deposited: | 06 Jan 2016 07:03 |
Last Modified: | 06 Jan 2016 07:03 |
URI: | http://eprints.iisc.ac.in/id/eprint/53053 |
Actions (login required)
View Item |