Kondiparthi, Mahesh (2010) Simultaneous measurement of multiple parameters using a single fiber Bragg grating. In: Conference on Infrared Remote Sensing and Instrumentation XVIII, AUG 01-03, 2010, San Diego, CA, .
PDF
7808-63.pdf - Published Version Restricted to Registered users only Download (563kB) | Request a copy |
Abstract
A novel approach for simultaneous measurement of chirp (any parameter that can induce strain gradient on FBG) and temperature using a single FBG is proposed. Change in reflectivity at central wavelength of FBG reflection & Bragg wavelength shifts induced due to temperature were used for chirp & temperature measurements respectively. Theoretical resolution limit for chirp and temperature using an Optical Spectrum Analyzer (OSA) with 1pm wavelength resolution and >58dB dynamic range are 12.8fm and 1/13 degrees C respectively.
Item Type: | Conference Paper |
---|---|
Series.: | Proceedings of SPIE-The International Society for Optical Engineering |
Publisher: | The International Society for Optical Engineering |
Additional Information: | Copyright of this article belongs to The International Society for Optical Engineering. |
Keywords: | simultaneous measurement; fiber Bragg grating; optical spectrum analyzer; reflectivity at central wavelength |
Department/Centre: | Division of Physical & Mathematical Sciences > Instrumentation Appiled Physics Division of Physical & Mathematical Sciences > Physics |
Date Deposited: | 07 Mar 2011 06:07 |
Last Modified: | 07 Apr 2011 07:00 |
URI: | http://eprints.iisc.ac.in/id/eprint/35914 |
Actions (login required)
View Item |