ePrints@IIScePrints@IISc Home | About | Browse | Latest Additions | Advanced Search | Contact | Help

Implementing Machine Learning approaches for accelerated prediction of bone strain in acetabulum of a hip joint

Nimmal Haribabu, G and Basu, B (2024) Implementing Machine Learning approaches for accelerated prediction of bone strain in acetabulum of a hip joint. In: Journal of the Mechanical Behavior of Biomedical Materials, 153 .

[img] PDF
jou_mec_beh_bio_mat_153_2024.pdf - Published Version
Restricted to Registered users only

Download (9MB)
Official URL: https://doi.org/10.1016/j.jmbbm.2024.106495

Abstract

The Finite Element (FE) methods for biomechanical analysis involving implant design and subject parameters for musculoskeletal applications are extensively reported in literature. Such an approach is manually intensive and computationally expensive with longer simulations times. Although Artificial Intelligence (AI) based approaches are implemented to a limited extent in biomechanics, such approaches to predict bone strain in acetabulum of a hip joint, are hardly explored. In this context, the primary objective of this paper is to evaluate machine learning (ML) models in tandem with high-fidelity FEA data for the accelerated prediction of the biomechanical response in the acetabulum of the human hip joint, during the walking gait. The parameters used in the FEA study included the subject weight, number and distribution of fins on the periphery of the acetabular shell, bone condition and phases of the gait cycle. The biomechanical response has also been evaluated using three different acetabular liners, including pre-clinically validated HDPE-20 HA-20 Al2O3, highly-crosslinked ultrahigh molecular weight polyethylene (HC-UHMWPE) and ZrO2-toughened Al2O3 (ZTA). Such parametric variation in FEA analysis, involving 26 variables and a full factorial design resulted in 10,752 datasets for spatially varying bone strains. The bone condition, as opposed to subject weight, was found to play a statistically significant role in determining the strain response in the periprosthetic bone of the acetabulum. While utilising hyperparameter tuning, K-fold cross validation and statistical learning approaches, a number of ML models were trained on the FEA dataset, and the Random Forest model performed the best with a coefficient of determination (R2) value of 0.99/0.97 and Root Mean Square Error (RMSE) of 0.02/0.01 on the training/test dataset. Taken together, this study establishes the potential of ML approach as a fast surrogate of FEA for implant biomechanics analysis, in less than a minute. © 2024 Elsevier Ltd

Item Type: Journal Article
Publication: Journal of the Mechanical Behavior of Biomedical Materials
Publisher: Elsevier Ltd
Additional Information: The copyright for this article belongs to the Elsevier Ltd.
Keywords: Alumina; Aluminum oxide; Biomechanics; Forecasting; Forestry; Fracture; High density polyethylenes; Hip prostheses; Machine learning; Mean square error; Nonmetallic matrix composites; Strain; Ultrahigh molecular weight polyethylenes; Zirconia, Acetabulum; Biomechanical analysis; Biomechanical response; Bone strains; Condition; Hip joints; Machine learning approaches; Machine learning models; Machine-learning; Subject weights, Finite element method
Department/Centre: Division of Chemical Sciences > Materials Research Centre
Date Deposited: 23 Apr 2024 11:15
Last Modified: 23 Apr 2024 11:15
URI: https://eprints.iisc.ac.in/id/eprint/84610

Actions (login required)