Reversal Trend of Hounsfield Unit Values of Substances with High and Low Effective Atomic Numbers

Zarei, F and Chatterjee, S and Chatterjee, VV and Haghighi, RR (2020) Reversal Trend of Hounsfield Unit Values of Substances with High and Low Effective Atomic Numbers. In: Iranian Journal of Medical Physics, 17 (5). pp. 340-349.

[img]
Preview
PDF
ira_jou_med_phy_17_5_340-349.pdf - Published Version

Download (1MB) | Preview
Official URL: https://dx.doi.org/10.22038/ijmp.2019.42841.1642.

Abstract

Introduction: In dual-energy computed tomography (DECT), the Hounsfield values of a substance measured at two different energies are the basic data for finding the chemical properties of a substance. The trends of Hounsfield unit (HU) alterations following the changes in energy are different between the materials with high and low Zeff. The present study aimed to analyze the basic principles related to the attenuation coefficient of x-ray photons and a quantitative explanation is given for the mentioned behavior or trend. Material and Methods: A mathematical expression was derived for the HU difference between two different scanner voltages. Attenuation coefficients of diverse substances, such as methanol, glycerol, acetic acid, the aqueous solution of potassium hydroxide, and water were calculated for x-ray scanners operating differently at distinct applied voltages and with diverse inherent or added filters. Results: Findings of the current study demonstrated that the negative or positive outcome of HU(V1) - HU(V2) equation is not determined by the electron density of a substance. However, it is affected by the effective atomic number (Zeff) of the material and machine parameters specified by the source spectrum. Conclusion: According to our results, the sign of HU difference HU(V1) � HU(V2) for the variable cases of V2 and V1 gives an indication of the effective atomic number of the material under study. The obtained results might be of diagnostic value in the DECT technique. © xxxx **

Item Type: Journal Article
Publication: Iranian Journal of Medical Physics
Publisher: Mashhad University of Medical Sciences
Additional Information: Copyright to this article belongs to Mashhad University of Medical Sciences
Keywords: Computerized tomography; Potassium hydroxide; Scanning; X rays, Applied voltages; Attenuation coefficient; Dual energy computed tomography (DECT); Effective atomic number; Hounsfield units; Hounsfield values; Machine parameters; Mathematical expressions, Atoms
Department/Centre: Division of Physical & Mathematical Sciences > Instrumentation Appiled Physics
Date Deposited: 24 Dec 2020 10:59
Last Modified: 24 Dec 2020 10:59
URI: http://eprints.iisc.ac.in/id/eprint/67241