ePrints@IIScePrints@IISc Home | About | Browse | Latest Additions | Advanced Search | Contact | Help

Comparison of Envelope Detection and Signal Normalization Methods for Foetal Heart Rate Extraction from Foetal Heart Sound

Amrutha, B and Arora, Manish (2018) Comparison of Envelope Detection and Signal Normalization Methods for Foetal Heart Rate Extraction from Foetal Heart Sound. In: Fourth International Conference on Biosignals, Images and Instrumentation (ICBSII), Chennai, India, 2018, 22-24 March 2018, Chennai, India.

[img]
Preview
PDF
eprints_icbsii.pdf - Accepted Version

Download (807kB) | Preview
Official URL: https://doi.org/10.1109/ICBSII.2018.8524718

Abstract

Foetal heart rate is an important indicator of foetal well-being which indicates fetal distress during antepartum and intrapartum period. This paper provides systematic comparison of four different methods envelope detection and four different methods of signal normalization to determine the fetal heart rate from fetal heart sound. To obtain the ground truth, each heart beat was manually identified by visualizing the sound recording on a computer screen. Four different measure of error are used compare various combinations of envelop detection and normalization methods. It is found that the homomorphic filtering gave highest accuracy. Choice of normalization methods did not affect accuracy measurement in most cases.

Item Type: Conference Paper
Publisher: IEEE
Additional Information: Copyright to this article belongs to IEEE.
Keywords: Correlation;Measurement uncertainty;Band-pass filters;Instruments;Fetal heart rate;fetal heart sound;envelope.
Department/Centre: Division of Mechanical Sciences > Centre for Product Design & Manufacturing
Date Deposited: 19 Jan 2019 17:06
Last Modified: 19 Jan 2019 17:06
URI: http://eprints.iisc.ac.in/id/eprint/61556

Actions (login required)

View Item View Item