Banerjee, Ayan and Rapol, Umakant D and Wasan, Ajay and Natarajan, Vasant (2001) High-accuracy wavemeter based on a stabilized diode laser. In: Applied Physics Letters, 79 (14). pp. 2139-2141.
![]()
|
PDF
High_accuracy.pdf Download (198kB) |
Abstract
We have built a high-accuracy wavelength meter for tunable lasers using a scanning Michelson interferometer and a reference laser of known wavelength. The reference laser is a frequency-stabilized diode laser locked to an atomic transition in Rb. The wavemeter has a statistical error per measurement of 5 parts in $10^7$, which can be reduced considerably by averaging. Using a second stabilized diode laser, we have verified that systematic errors are below 4 parts in $10^8$.