A Simple Instrumentation Calibration Technique for Electrical Impedance Tomography (EIT) Using A 16-Electrode Phantom

Bera, Tushar Kanti and Nagaraju, J (2009) A Simple Instrumentation Calibration Technique for Electrical Impedance Tomography (EIT) Using A 16-Electrode Phantom. In: IEEE International Conference on Automation Science and Engineering, AUG 22-25, 2009, Bangalore, pp. 347-352.

[img] PDF
simple.pdf - Published Version
Restricted to Registered users only

Download (1MB) | Request a copy
Official URL: http://ieeexplore.ieee.org/search/srchabstract.jsp...

Abstract

A simple analog instrumentation for Electrical Impedance Tomography is developed and calibrated using the practical phantoms. A constant current injector consisting of a modified Howland voltage controlled current source fed by a voltage controlled oscillator is developed to inject a constant current to the phantom boundary. An instrumentation amplifier, 50 Hz notch filter and a narrow band pass filter are developed and used for signal conditioning. Practical biological phantoms are developed and the forward problem is studied to calibrate the EIT-instrumentation. An array of sixteen stainless steel electrodes is developed and placed inside the phantom tank filled with KCl solution. 1 mA, 50 kHz sinusoidal current is injected at the phantom boundary using adjacent current injection protocol. The differential potentials developed at the voltage electrodes are measured for sixteen current injections. Differential voltage signal is passed through an instrumentation amplifier and a filtering block and measured by a digital multimeter. A forward solver is developed using Finite Element Method in MATLAB7.0 for solving the EIT governing equation. Differential potentials are numerically calculated using the forward solver with a simulated current and bathing solution conductivity. Measured potential data is compared with the differential potentials calculated for calibrating the instrumentation to acquire the voltage data suitable for better image reconstruction.

Item Type: Conference Paper
Publisher: IEEE
Additional Information: Copyright 2009 IEEE. Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works must be obtained from the IEEE.
Keywords: analog instrumentation; electrical impedance tomography; calibration; practical biological phantom; forward solver; finite element method
Department/Centre: Division of Physical & Mathematical Sciences > Instrumentation Appiled Physics
Date Deposited: 09 Jun 2010 05:18
Last Modified: 19 Sep 2010 06:00
URI: http://eprints.iisc.ac.in/id/eprint/27265