Sinha, Shantanu and Srinivasan, R (1983) High-Pressure Low-Temperature Locknut Cell For Both Electron-Paramagnetic-Res And Nmr-Studies To 10 Kilobars And 77-K. In: Review of Scientific Instruments, 54 (11). pp. 1492-1496.
PDF
GetPDFServlet.pdf - Published Version Restricted to Registered users only Download (462kB) | Request a copy |
Abstract
A locked high-pressure cell with working pressure range up to 10 kbars suitable for low-temperature studies to 77 K has been described. It can be used for both EPR and NMR studies of single crystals (and other solid samples). The high-pressure seal and all other aspects of the cell remain the same for either application. Only a change of the bottom plug is required for a switch from a nuclear-magnetic-resonance (NMR) to an electron-paramagnetic-resonance (EPR) experiment. Details of the procedure for the calibration of pressure inside the cell at various temperatures are discussed. The performance of the cell in EPR (Cr3+ion) and NMR (27Al nucleus) studies is reported.
Item Type: | Journal Article |
---|---|
Publication: | Review of Scientific Instruments |
Publisher: | American Institure of Physics |
Additional Information: | Copyright of this article belongs to American Institure of Physics. |
Department/Centre: | Division of Physical & Mathematical Sciences > Physics |
Date Deposited: | 12 Aug 2009 06:10 |
Last Modified: | 19 Sep 2010 05:32 |
URI: | http://eprints.iisc.ac.in/id/eprint/20271 |
Actions (login required)
View Item |