Nagaboopathy, M and Vijayanand, C and Hegde, Gopalkrishna and Reddy, KPJ and Arunan, E (2008) Single-pulse chemical shock tube for ignition delay measurements. In: Current Science, 95 (1). pp. 78-82.
|
PDF
sdarticle.pdf Download (163kB) |
Abstract
We describe a single-pulse chemical shock tube CST2 established for measuring the reaction rate of chemical reactions and ignition delay for fuels at high temperature along with the procedure for its calibration. The suitability of the facility for measuring the ignition delay is demonstrated by measuring the ignition delay for the ethane–oxygen gas mixture in the temperature range 1250–1611 K by recording the ignition-induced pressure jump and emission from CH radical simultaneously.The results obtained in the present study compare well with the earlier reported values.
Item Type: | Journal Article |
---|---|
Publication: | Current Science |
Publisher: | Indian Academy of Sciences |
Additional Information: | Copyright of this article belongs to Indian Academy of Sciences. |
Keywords: | Chemical kinetics;ignition delay;reaction rate;shock tube. |
Department/Centre: | Division of Mechanical Sciences > Aerospace Engineering(Formerly Aeronautical Engineering) Division of Chemical Sciences > Inorganic & Physical Chemistry |
Date Deposited: | 06 Aug 2008 |
Last Modified: | 19 Sep 2010 04:48 |
URI: | http://eprints.iisc.ac.in/id/eprint/15477 |