Murthya, OVSN and Venkataraman, V (2007) Construction and calibration of a 12 T pulsed magnet integrated with a 4 K closed-cycle refrigerator. In: Review of Scientific Instruments, 78 (11). 113905-1-113905-5.
PDF
GetPDFServlet.pdf Restricted to Registered users only Download (158kB) | Request a copy |
Abstract
A low cost 12 T pulsed magnet system has been integrated with a closed-cycle helium refrigerator for performing magnetotransport measurements. Minimal delay between pulses and ac current excitation with software lock-in to reduce noise enables quick but accurate measurements to be performed at temperatures of 4–300 K up to 12 T. An additional pulsed magnet operating with a liquid nitrogen cryostat extends the range up to 19 T. The instrument has been calibrated against a commercial superconducting magnet by comparing quantum Hall effect data in a p-channel SiGe/ Si heterostructure, and common issues arising out of pulsed magnet usage have been addressed. The versatility of the system is demonstrated through magnetotransport measurements in a variety of samples such as heterostructures, narrow gap semiconductors, and those exhibiting giant magnetoresistance.
Item Type: | Journal Article |
---|---|
Publication: | Review of Scientific Instruments |
Publisher: | American Institute of Physics |
Additional Information: | Copyright of this article belongs to American Institute of Physics. |
Department/Centre: | Division of Physical & Mathematical Sciences > Physics |
Date Deposited: | 28 May 2008 |
Last Modified: | 19 Sep 2010 04:45 |
URI: | http://eprints.iisc.ac.in/id/eprint/14103 |
Actions (login required)
View Item |