Madhu, BR and Narayanan, PS (1983) On the factors affecting the performance of a continuous flow cryostat. In: Indian Journal of Cyrogenics, 8 (2). pp. 91-94.
Full text not available from this repository. (Request a copy)Abstract
A continuous flow cryostat was designed (80-300 K) for the study of dielec. and thermal expansion behavior of solid samples at high vacuum (.apprx.10-5 torr). The temp. up to 150 K was reached by increasing the opening or a control valve near the suction pump. Further opening of the valve for obtaining lower temps. revealed a slow oscillation of temp. between 150 and 125 K. And several methods tried out for avoiding the temp. oscillations are discussed