Varadharaj, Eswaramoorthy K and Jampana, Nagaraju (2015) Development and Evaluation of Prussian Blue Mediated Glucose Sensor for Reverse Iontophoresis Application. In: ECS ELECTROCHEMISTRY LETTERS, 4 (11). B13-B16.
Full text not available from this repository. (Request a copy)Abstract
The screen printed electrochemical glucose sensor is developed suitable for revere iontophoresis (RI) application. Glucose oxidase is immobilized on screen printed sensor using crosslinking method. Electrochemical and material characterization studies are conducted on the developed sensor and the obtained results confirm the suitability of the developed sensor for RI application. The developed sensor is validated by conducting clinical investigations on 10 human subjects through RI. A correlation is established between the blood glucose and extracted glucose, and correlation is found to be 0.73. (C) 2015 The Electrochemical Society. All rights reserved.
Item Type: | Journal Article |
---|---|
Publication: | ECS ELECTROCHEMISTRY LETTERS |
Publisher: | ELECTROCHEMICAL SOC INC |
Additional Information: | Copy right for this article belongs to the ELECTROCHEMICAL SOC INC, 65 SOUTH MAIN STREET, PENNINGTON, NJ 08534 USA |
Department/Centre: | Division of Physical & Mathematical Sciences > Instrumentation Appiled Physics |
Date Deposited: | 03 Dec 2015 04:52 |
Last Modified: | 03 Dec 2015 04:52 |
URI: | http://eprints.iisc.ac.in/id/eprint/52854 |
Actions (login required)
View Item |