Development of a microstrip gas chamber to study the effect of drift gap on its performance

Radhakrishna, V and Rajanna, K and Desai, SS and Shaikh, AM (2001) Development of a microstrip gas chamber to study the effect of drift gap on its performance. In: Review of Scientific Instruments, 72 (2). pp. 1361-1365.

[img] PDF
Development_of_a_microstrip.pdf
Restricted to Registered users only

Download (88kB) | Request a copy

Abstract

We present a new microstrip gas chamber design and a detailed experimental study of the effect of drift gap on its performance. The chamber has the provision to vary the drift gap in situ which enables one to carry out the performance study without altering the other set operating parameters such as gas pressure, anode voltage, and drift voltage. The effect of drift gap on detector gain and energy resolution is discussed here. The importance of optimization of drift gap is highlighted. The maximum gain obtained with P10 $(Ar90\%{+}CH_410\%)$ gas is $\sim750$ and the best resolution achieved is $\sim19\%$ for 5.9 keV x rays.

Item Type: Journal Article
Publication: Review of Scientific Instruments
Publisher: American Institute of Physics
Additional Information: Copyright of this article belongs to American Institute of Physics.
Department/Centre: Division of Physical & Mathematical Sciences > Instrumentation Appiled Physics
Date Deposited: 09 Feb 2007
Last Modified: 19 Sep 2010 04:28
URI: http://eprints.iisc.ac.in/id/eprint/7451