ePrints@IIScePrints@IISc Home | About | Browse | Latest Additions | Advanced Search | Contact | Help

Evaluation of implant properties, safety profile and clinical efficacy of patient-specific acrylic prosthesis in cranioplasty using 3D binderjet printed cranium model: A pilot study

Basu, B and Bhaskar, N and Barui, S and Sharma, V and Das, S and Govindarajan, N and Hegde, P and Perikal, PJ and Antharasanahalli Shivakumar, M and Khanapure, K and Tekkatte Jagannatha, A (2021) Evaluation of implant properties, safety profile and clinical efficacy of patient-specific acrylic prosthesis in cranioplasty using 3D binderjet printed cranium model: A pilot study. In: Journal of Clinical Neuroscience, 85 . pp. 132-142.

[img] PDF
jou_cli_neu_85_132-142_2021.pdf - Published Version
Restricted to Registered users only

Download (3MB) | Request a copy
Official URL: https://dx.doi.org/10.1016/j.jocn.2020.12.020

Abstract

There exists a significant demand to develop patient-specific prosthesis in reconstruction of cranial vaults after decompressive craniectomy. we report here, the outcomes of an unicentric pilot study on acrylic cranial prosthesis fabricated using a 3D printed cranium model with its clinically relevant mechanical properties. Methods: The semi-crystalline polymethyl methacrylate (PMMA) implants, shaped to the cranial defects of 3D printed cranium model, were implanted in 10 patients (mean age, 40.8 ± 14.8 years). A binderjet 3D printer was used to create patient-specific mould and PMMA was casted to fabricate prosthesis which was analyzed for microstructure and properties. Patients were followed up for allergy, infection and cosmesis for a period of 6 months. Results: As-cast PMMA flap exhibited hardness of 15.8 ± 0.24Hv, tensile strength of 30.7 ± 3.9 MPa and elastic modulus of 1.5 ± 0.1 GPa. 3D microstructure of the semi-crystalline acrylic implant revealed 2.5�15 µm spherical isolated pores. The mean area of the calvarial defect in craniectomy patients was 94.7 ± 17.4 cm2. We achieved a cranial index of symmetry (CIS -) of 94.5 ± 3.9, while the average post-operative Glasgow outcome scale (GOS) score recorded was 4.2 ± 0.9. Conclusions: 3D printing based patient-specific design and fabrication of acrylic cranioplasty implant is safe and achieves acceptable cosmetic and clinical outcomes in patients with decompressive craniectomy. Our study ensured clinically acceptable structural and mechanical properties of implanted PMMA, suggesting that a low cost 3D printer based PMMA flap is an affordable option for cranioplasty in resource constrained settings. © 2021 Elsevier Ltd

Item Type: Journal Article
Publication: Journal of Clinical Neuroscience
Publisher: Churchill Livingstone
Additional Information: The copyright of this article belongs to Churchill Livingstone
Department/Centre: Division of Chemical Sciences > Materials Research Centre
Division of Interdisciplinary Sciences > Centre for Biosystems Science and Engineering
Date Deposited: 22 Feb 2021 06:26
Last Modified: 22 Feb 2021 06:26